Multi-Drug Oral Fluid Test 3-in-1

Multi-Drug Oral Fluid Test 3-in-1


  • Brand:Testera
  • Product Code:IVD-MD01
  • Ships in:1-2 days

  • $8.20


  • Simple test procedure
  • Easy visual interpretation
  • Rapid detection results in 5 minutes
File Size
Multi-drug-3-drugs.pdf 2.04MB Download

The Multi-Drug Rapid Test is a lateral flow chromatographic immunoassay for the qualitative detection of multiple drugs or drug metabolites in human oral fluid at the following cut-off concentrations:

  • Marijuana (THC)50 ng/mL
  • Cotinine (COT) 20 ng/mL
  • Synthetic Marijuana 25 ng/mL

The Multi-Drug Rapid Test is a rapid screening test that can be performed without the use of an instrument. The test utilizes monoclonal antibodies to selectively detect elevated levels of specific drugs in specimen.

Testing Principles

Allow the test device, specimen and/or controls to reach room temperature (15-30 °C) prior to testing. Instruct the donor to not place anything in the mouth including food, drink, gum or tobacco products for at least 10 minutes prior to collection.

  1. 1. Bring the pouch to room temperature before opening it. Remove the test from the sealed pouch and use it within one hour.
  2. 2. Take off the Device cap and collect oral fluid specimen as follows.
    Important: Place the absorbent wick against the upper, lower jaws and roots to enrich theoral fluid. Insert the sponge end into the mouth, actively swab around the gums on both sides of the mouth (10-15 times) to assist saturation. Put the absorbent wick under the tongue to collect oral fluid until the flow appear in the test windows (approximately 60 seconds) and then take out the device and start a timer.
    If no flow appeared repeat the procedure in steps above until the flow appear.
    If no flow appeared after triplicate of steps above, discard the device, review procedures with the donor and repeat the test using a new device.
  3. 3. Place the test device on a clean and level surface.
  4. 4. Read the test result at 3-10 minutes.

If all lines are clearly visible at 3 minutes or sooner, then the test can be interpreted as negative and discarded. If any lines are not visible at 3 minutes, then the test should be re-read at 10 minutes

Expiration Date: 12/2025